RephiLe Bioscience Large Direct-Pure EDI Water Systems
The large Direct-Pure® EDI system produces Type II water from tap water directly. Resistivity of product water is above 5 MΩ·cm (25oC) ,which meets or exceeds Type Ⅱ water quality standard defined by ASTM®, CAP, CLSI and ISO® 3696/BS 3997, the European and U.S. Pharmacopoeia. The System is designed for users who need 400 to 1800 liters of Type II water per day.
Features
• The system incorporates the best in class Siemens EDI module to deliver high quality Type II water consistently at relatively lower cost.
• Automatic water quality control programs ensure EDI product water meeting or exceeding Type II water requirements.
• The fully automated control system contains pre-rinse, RO membrane fouling preventing flush, RO membrane life monitoring as well as many other programs for convenient operation.
• The pre-rinsing program starts automatically when a new RO membrane is installed so there is no need for users to set up additional cleaning procedures.
• Users can set water dispensing time for easy water collection.
• RephiLe can provide full document support to meet users’ GMP, GLP, FDA and other certification requirements.
Main Applications
Type II Pure water can be used in many areas. Some of the typical applications are.
• Preparation of bio-reagents
• Preparation of culture media
• Preparation of solution for chemical analysis such as HPLC, ICP etc.
• For clinical analyzers
• Feed water for ultrapure water systems
• Medical device and equipment rinsing
• For serum and blood fractions
• For ophthalmics
Specifications
Large Direct-Pure® Water System, EDI |
|
Feed water quality |
|
Tap water* |
Municipal water, TDS <1000 ppm (2000 μS/cm) |
Water temperature |
5~45 oC |
Water pressure |
30~90 psi (2-6 bar) |
Product Water Flow Rate Typically 30, 60 or 125 L/h (at 25 oC) |
|
Product Water (Type II ) Quality |
|
Resistivity |
>5 MΩ.cm @ 25 oC (typically > 10 MΩ.cm @ 25°C) |
TOC |
< 30 ppb |
Particles in Ultrapure Water (>0.2 μm) |
< 1 /mL (with a 0.2μm final filter) |
Microorganisms |
< 1 cfu/ml (with a 0.2μm final filter) |
Pyrogens (endotoxins) |
< 0.001Eu/ml (with a terminal ultrafiltration cartridge) |
Dimensions |
|
Length x Depth x Height |
37 cm x 61 cm x 110cm |
* If water hardness is greater than 1000 ppm, RephiLe suggests using an additional water softening device
in order to protect the RO membrane in your system.
Ordering Information
Description |
Catalogue No. |
Direct-Pure Water System, EDI 30 |
RD0E03000 |
Direct-Pure Water System, EDI 60 |
RD0E06000 |
Direct-Pure Water System, EDI 125 |
RD0E01H00 |