RephiLe Bioscience Direct-Pure UP20 Ultra Pure & RO Water System
osmosis (RO) water from tap water directly. Quality of ultrapure water produced meets or exceeds ASTM, CLSI, CAP, and ISO Type I water standards.
Features
• Electrical and water compartments are completely segregated, thus safer to operate.
• The fully automated control system contains pre-rinse, RO membrane fouling preventing flush, RO membrane
and cartridge life monitoring as well as many other programs for convenient operation.
• The pre-rinsing program starts automatically when a new RO membrane is installed so there is no need for users
to set up additional cleaning procedures.
• Two sets of dual-column purification cartridges work sequentially to ensure thorough removal of ions and organics
in water.
• The system recirculates automatically every hour to prevent microbial from growing.
• Resistivity is measured by a high-precision resistivity meter.
• Users can set water dispensing time for easy water collection.
• The system is equipped with a 30-liter PE tank(upgradable to a 60-liter) and a liquid level sensor,
• A 0.2 μm final filter or ultrafiltration filter (optional) removes final contaminants.
• RephiLe can provide full document support to meet users’ GMP, GLP, FDA and other certification requirements.
Main Applications
Pure water can be used in many areas. Here are some typical applications:
Ultrapure Water |
RO Pure Water |
Applications |
Applications |
• HPLC mobile phase preparation • Preparation of reagent blank solution • As sample diluents for GC, HPLC, AA, ICP-MS and other analytical techniques • Preparation of buffer and culture media for mammalian cell culture • Preparation of molecular biology reagents, etc. |
• Glassware cleaning • Washing machine for glassware • Water bath water • Autoclave . |
Specifications
Feed water quality |
|
Tap water* |
Municipal water, TDS <1000 ppm (2000 μS/cm) |
Operating temperature |
5~45 °C |
Feed water pressure |
15~60 psi (1.0~4.0 bar) |
Product Water Flow Rate |
|
Ultrapure water (Type I) |
Typically 1.5 L/min (at 25°C ) |
RO water |
Typically 20 L/h (at 25°C) |
Product Water Quality |
|
Resistivity |
>18.2 MΩ.cm at 25°C |
TOC |
<10ppb (<5 ppb with a dual wavelength UV lamp)** |
Particles in Ultrapure Water (>0.2 μm) |
< 1 /mL (with a 0.2µm final filter) |
Microorganisms |
< 1 cfu/ml (with a 0.2µm final filter) |
Pyrogens (endotoxins) |
< 0.001Eu/ml (with a terminal ultrafiltration cartridge) |
Dimensions |
|
Length x Depth x Height |
30 cm x 49 cm x 48 cm |
Net Weight (Operating Weight) |
22~25 kg (25~32 kg) |
* If water hardness is greater than 1000 ppm, RephiLe suggests using an additional water softening device in order
to protect the RO membrane in your system.
** Product water quality may vary due to local feedwater conditions.
Ordering Information
Description |
Catalogue No |
Direct-Pure System |
|
Direct-Pure UP 20 Ultrapure and RO Water System |
RD0P02000 |
Consumables |
Catalogue No. |
RephiDuo H Pack cartridge |
RR300Q201 |
RephiDuo U Pack cartridge |
RR600Q201 |
RephiDuo U Pack cartridge with a 0.2 μm PES Capsule Filter |
RRQ2F6SKT |
RO membrane Direct-Pure UP 20 System, 1/PK |
RAR010001*2 |
Polypropylene Depth Filter, 10 Inches 1 µm, 12/Pk |
LABPA0112 |
Polypropylene Depth Filter, 10 Inches 10 µm, 12/Pk |
LABPA1012 |
Activated Carbon cartridge, 10 Inches, 12/Pk |
LABAC1012 |
Softening carridge, 10 Inches, 12/Pk |
LABAT1012 |
0.2 μm PES Final Capsule Filter |
RAFFC0250 |
30-liter PE tank |
RATANK030 |
60-liter PE tank |
RATANK060 |